In electrophysiological experiments, even very small changes in voltage or current—occurring on millisecond or microsecond timescales—can reveal critical neuronal signaling mechanisms. Capturing these subtle electrical events requires highly sensitive equipment designed for precision measurement.
However, experimental sensitivity can often be compromised by the noise, not by inherent instrument noise, but by interference from external electromagnetic fields. Common sources of such interference include AC power lines (50/60 Hz), fluorescent and LED lighting systems, and electromagnetic emissions from nearby electronic devices. These external signals are frequently introduced due to suboptimal wiring practices within electrophysiology rigs.
This article covers best practices for grounding, shielding, and headstage REF/GND connection to minimize external interference pickup.
A recommended practice for grounding electrophysiology apparatus is the star grounding technique. This involves connecting all equipment grounds to a single, common grounding point, typically called the “star” node. From this central node, a single grounding conductor is connected to a dedicated earth ground or grounded wall outlet.
Each XDAQ includes chassis ground and system ground ports on the rear panel by the power input. Here’s how they differ:
The two-port design gives users flexibility while enabling a clean single-point ground when bonded together.
You may use the ground port on the XDAQ as the common grounding point for all apparatus in your electrophysiology rig. We recommend star grounding to avoid ground-loop formation and electric-field interference—run a dedicated ground lead from each instrument to the common point (the XDAQ ground port).
Tips:
Users can try shorting (bonding) these two ports and then connecting the Faraday cage and rest of the instrumentation within the rig to that bonded point—this arrangement has eliminated most noise pickup in our lab environment.
Effective shielding is crucial for reducing interference from external electric fields, particularly those originating from mains power lines. Although notch filters can attenuate the fundamental 50/60 Hz mains frequency, higher-order harmonics often persist and may cause analog-to-digital converter (ADC) saturation and signal distortion.
A Faraday cage—typically constructed from conductive materials—can block these electromagnetic disturbances by enclosing the experimental apparatus and redirecting external electric fields. When integrated properly, a Faraday cage enables more accurate and stable signal acquisition.
We recommend conducting electrophysiological recordings within a Faraday cage to minimize electromagnetic interference.
If you are constructing a faraday cage yourself, here are a few suggestions:
XDAQ is designed with isolated headstage ports, which means the electrical circuits at the headstages (connected via the headstage ports) are protected, or isolated from that of the XDAQ’s digital system. This minimizes any electrical noise propagating from the XDAQ and its connected instrumentations. Properly connected REF/GND signal between the headstage to the electrode and the animal's reference is essential for ensuring signal integrity and minimizing noise.
Suggestions for connecting the headstage REF/ and GND to the animal:
Info:
For establishing a stable REF/GND point at the animal site, we recommend experiment with various headstage GND/REF configuration and locations for your prep and always verify reliable signal is supplied to the headstage. Shorting GND and References together is a common practice and could be a good starting point. Once gaining a base baseline recording quality from a shorted GND-REF configuration, users could consider experimenting with separating these signals. For example, one might use the metal canular from the probe as REF, as it is close to the recording sites, and use a bone screw as a GND. Some neural probes have dedicated electrode sites that are designed for use as REF.
Tips:
When adapters are used, check the entire signal path for GND and REF between the electrode and the headstage. Check out v2 KonteX headstage adapters for added flexibility in GND/REF configuration.